13-26
IM 701610-01E
13.11 Setting the Time Difference from the GMT
(Greenwich Mean Time) and SNTP
Relevant Keys
ACTION DELAY
X - Y
MENU MENU
PHASE
SEARCH
HORIZONTAL
VERTICAL
TRIGGER
TRIG D
HELP
ESC
SELECTRESET
SIMPLE
MODE POSITION
MATH
MISCFILE
CLEAR
TRACE
HISTORY
MEASURE
SETUP
PRESET
DISPLAY
CURSOR
CH
1
CH
2
CH
3
CH
4
POWER
COPY
IMAGE SAVE
SHIFT
ZOOM
GO/NO-GO
ENHANCED
ACQSTART/STOP
SNAP
SHOT
VDIV TIME DIV
Procedure
1. Press MISC.
2. Press the Network soft key to display the Network menu.
3. Press the Others soft key to display the Others settings menu.
4. Turn the jog shuttle to select Time Hour for Time Difference From GMT.
Settings for the following items can be
input directly using a USB keyboard. ( )
5. Press SELECT to display a dialog box where you can make time settings.
6. Turn the jog shuttle to set the number of hours in the range of -12 to 13.
7. Likewise, set the Minute of Time Difference From GMT in the range of 0 to 59.
Note
Minute can be specified only if Time Hour is set in the range of –11 to 12. (Minute cannot be
specified if Time Hour is set to –12 or 13.)
Setting Up SNTP (Simple Network Time Protocol, Version 1.13 or Later)
8. Turn the jog shuttle to select Exec at Power On.
9. Select ON or OFF by pressing SELECT.
10. Turn the jog shuttle to select Server.
11. Press SELECT to display the keyboard.
12. Enter the IP address for the SNTP server.
13. Turn the jog shuttle to select Time out.
14. Press SELECT to display the time settings box.
15. Turn the jog shuttle to set the time out time in the range from 1-60.
Commenti su questo manuale