Axis BTA 2100D Manuale dell'Operatore Pagina 248

  • Scaricare
  • Aggiungi ai miei manuali
  • Stampa
  • Pagina
    / 597
  • Indice
  • SEGNALIBRI
  • Valutato. / 5. Basato su recensioni clienti
Vedere la pagina 247
246
User Defined Settings
Item Description
Fill Volume (uL) 0 to 10.0 µL with 10 µL syringe
0 to 5.0 µL with 5 µL syringe
Specifies the sample volume that will be used for each fill stroke.
Fill Strokes 0 to 99
Specifies the number of the times the sample will be “pumped” in
and out of the syringe before loading the sample volume into the
syringe.
Sample Air Gap Yes or No
If Yes, 1 µL of room air will be drawn into the syringe to create
an air gap before the sample plug.
Air Plug after
Sample
0 to 10.0 µL with 10 µL syringe
0 to 5.0 µL with 5 µL syringe
Specifies the volume of room air that will be drawn into the
syringe after it is loaded with sample.
Viscosity Settings
Item Description
Viscosity Delay (sec) 0 to 9.9 seconds
Specifies how long the syringe is to remain in the sample
vial after drawing up the sample.
Fill Speed (uL/sec) 0.1 to 50.0 µL per second with 10 µL syringe
0.1 to 25.0 µL per second with 5 µL syringe
Specifies the speed at which the sample will be drawn up
into the syringe.
Vedere la pagina 247
1 2 ... 243 244 245 246 247 248 249 250 251 252 253 ... 596 597

Commenti su questo manuale

Nessun commento